A bench top cyanoacrylate fuming cabinet
for the safe and controlled development of
latent fingerprints
Developing latent fingerprints using the cyanoacrylate fuming process is now
faster and more convenient with Foster & Freeman's new bench top MVC™1000,
the latest and smallest fuming cabinet to join the MVC™ range. The new cabinet
retains all the benefits and safety features of the larger versions but has a
complete processing time of only 30 minutes and is easy installed anywhere.
Manual or automatic operation
The MVC™1000 provides both manual and automatic modes of operation. Under
manual control, each phase of the cyanoacrylate fuming process 'humidification',
'fuming' and 'fume extraction' is initiated separately, enabling the operator to
extend or reduce the fuming process as required. Under automatic control each
phase proceeds without operator intervention and according to preset parameters
(cyanoacrylate fume exposure can be preset by the operator to between 3 and 99
minutes). During the fuming process, fans continually circulate the atmosphere in
the chamber to ensure that fingerprint development is uniform while sensors
monitor and control the internal humidity to maintain optimum conditions.
Self-contained fume cleansing system
Fume extraction is accomplished via an internal, fan-driven, activated carbon filter
with the capacity to perform over 150 cleansing cycles per refill. The self-contained
system avoids the need for expensive fume ducting systems and enables the
cabinet to be re-sited with minimum disturbance.
Health and Safety features
The MVC™1000 is designed to provide a safe working environment, ensuring that
the concentration of cyanoacrylate vapour in the atmosphere outside the cabinet is
always below 0.3ppm, the recognised Occupational Exposure Standard for the
substance.
A potential hazard of any cyanoacrylate fuming cabinet arises from the
cyanoacrylate deposits that form on the internal surfaces of the cabinet. Outgassed
vapour from these deposits, if allowed to build up, becomes a health
hazard to the operator, exposing him to a potentially harmful concentration of
cyanoacrylate vapour as soon as the door is opened. In the MVC™1000 this
hazard is eliminated as the fume cleansing system is activated automatically for a
short period every ten minutes to ensure the out-gassed vapours are removed.
To download the MVC™1000 product brochure please click on the following link: